Send Message

Guangdong Hongtuo Instrument Technology Co.,Ltd

Home ProductsPlastic Testing Machine

Lab 0~1000 °C Needle Flame Test Chamber For Insulating Materials

Lab 0~1000 °C Needle Flame Test Chamber For Insulating Materials

  • Lab 0~1000 °C Needle Flame Test Chamber For Insulating Materials
Lab 0~1000 °C Needle Flame Test Chamber For Insulating Materials
Product Details:
Place of Origin: China
Brand Name: Dahometer
Certification: ISO
Model Number: HT-5169T-N
Payment & Shipping Terms:
Minimum Order Quantity: 1 Set
Price: negotiable
Packaging Details: Standard Export Wooden Case
Delivery Time: 7days
Payment Terms: T/T, Western Union
Supply Ability: 1000 Set Per Months
Contact Now
Detailed Product Description
Name: Lab 0~1000 °C Needle Flame Test Chamber For Insulating Materials Needle Burner: Inner Hole Φ 0.5mm ± 0.1mm
Application:: Lighting, Low-voltage Electrical Appliances Etc Keyword: Needle Flame Flammability TesterNeedle Flame Test Chamber
Standards:: IEC 60695-2-2: 1991,IEC 60695-2-2 (1994-03) Etc Temperature Test Range: 0~1000 °C
High Light:

Insulating Materials Needle Flame Test Apparatus

,

Lab Needle Flame Test Apparatus

,

Lab Needle Flame Test Equipment

Lab 0~1000 °C Needle Flame Test Chamber For Insulating Materials​

 

Model: HT-5169T-N

 

Application:

The electric leakage between different polar parts or between the live parts and the grounded metal parts may cause leakage of insulation, and the arc generated may cause breakdown of the electric appliance. Short circuit or material corrosion caused by discharge, and even ignition caused a fire. This test instrument is a destructive test for the insulation material to simulate the above conditions, to test and evaluate the relative tracking resistance of the insulator under the action of the electric field and the impurity-containing water under the specified voltage. The leakage tracking tester is suitable for electrical and electronic products, solid electrical insulation materials for household appliances and their products, such as relay sockets, transfer switch covers, contact body and so on.

 

 

Standards:

IEC 60695-2-2: 1991 (Second Edition), Amendment No. 1 of IEC 60695-2-2 (1994-03), IEC60695 -11-5, GB/T 5169.5-2008, GB4706.1-2005 The simulated test project specified by the standard.

 

Specification:

1. Needle burner: inner hole Φ 0.5mm ± 0.1mm, outer diameter ≤Φ 0.9mm, length ≥ 35mm, stainless steel

 

2. Burner angle: tilt 45 ° (test) or vertical (when adjusting the flame height)

 

3. Burning time: 0-999.9s ± 0.1s adjustable (usually 30s)

 

4. Holding time: 0-999.9s ± 0.1s, automatic recording, manual suspension

 

5. Flame height: 12mm ± 1mm (with height measuring instruments)

 

6. The gas body: 95% butane gas (can use bottled lighter gas)

 

7. Temperature test range: 0~1000 °C

 

8. Flame temperature requirements: from 100 ° C ± 2 ° C to 700 ° C ± 3 ° C time within 23.5 seconds ± 1 second

 

9. Temperature thermocouple: Φ0.5mm imported armored (K type) thermocouple

 

10. Needle flame burning tester test background: black background

 

11. Test process: automatic control of test procedures, independent ventilation

 

12. Reference standard: IEC60695 -11-5, GB/T5169.5-2008, GB4706.1-2005

 

13. The studio volume: 0.5 cubic (can be customized according to customer requirements 0.75 cubic or 1 cubic)

 

14. The outer dimensions of the equipment: 1000mm wide × 650mm deep × 1300mm high

 